Evolution Pro UV-Vis Spectrophotometers for Research and Quality Control

The Thermo Scientific Evolution Pro UV-Vis Spectrophotometer line integrates durable, compatible hardware with a broad selection of accessories, and practical, easy-to-use Thermo Scientific™ Insight™ Pro Software. The result is routine, complex research and advanced applications, including current compliance.

Product Overview

  • Workflow-focused application modules make it easier to customize even the most intricate processes.
  • Users may swiftly move from sample to final report, thanks to the user-friendly INSIGHT Pro Software, which supports every step of the workflow.
  • Users can achieve the findings they need for quantitative analysis, scanning, and kinetics applications, with extensive software capabilities for data collection, processing, and reporting.
  • By guaranteeing data security, optional INSIGHT Pro Security Software capabilities help pharmaceutical laboratories comply with current 21 CFR Part 11 regulations.

Enhance Productivity with Sampling Flexibility

  • Thermo Scientific Smart Accessories feature a snap-in design that ensures ease of use and consistency, enabling users to complete analysis more quickly.
  • Make use of the broad beam separation and big sample compartment of the Evolution Pro UV-Vis Spectrophotometer, which enables unique accessories and creative applications.
  • Manual setup is eliminated by software integration and automatic recognition of sippers, cell changers, and autosampler accessories.

Regulatory/Quality Control Environments

To help users use the Evolution Pro UV-Vis Spectrophotometer in regulated or quality-control environments while reducing installation time and preserving system performance, a variety of compliance solutions and services are available.

Evolution Pro Pharma Value Bundles Include

  • Evolution Pro Spectrometer
  • INSIGHT Pro with Security Software
  • Qualification Filter Kit for UV-Vis
  • Validator for Insight Pro US Ltr or INTL A4 Paper Version

Evolution Pro USP/EP Standards Pharma Bundles Include

  • USP and EP Visible Standards Set
  • Validator for Insight Pro US Ltr or INTL A4 Paper Version
  • Evolution Pro Spectrometer
  • INSIGHT Pro with Security Software
  • USP and EP UV Standards Set

Accessory Ordering Information

Each accessory can be ordered individually. Cell changers, cuvettes and test holders, calibration carousels, sippers, fiber optic probes, small-volume sample instruments, and accessories for quick mixing and reflectance are further choices.

Download the Brochure for More Information

Specifications

Source: Thermo Fisher Scientific - UV-Vis Spectroscopy

Specifications
Baseline Flatness ± 0.001 A (200 to 800 nm), 2 nm SBW, smoothed
Drift < 0.0005 A&fraslhr., 500 nm, 2 nm SBW, 2-hour warmup
Electrical Requirements 100/240 V, 50/60 Hz
For Use With (Application) Academic, Chemicals, Food and Beverage, Chemicals Environmental, Industrial QA/QC, Material Science, Pharmaceuticals
Green Features ACT Ecolabel (Gold)
Includes Evolution Pro UV-Vis Spectrophotometer
Lamp Life Five years or longer if not using a live signal
Noise 0 A: < 0.00018 A
1 A: < 0.00022 A
2 A: < 0.00050 A
500 nm, 2 nm SBW, RMS
Optical Design Modified Ebert
Double beam with sample and reference cuvette/accessory positions
Pharmacopoeia
Compliance Testing
Wavelength Accuracy: ± 0.3 nm 190–900 nm
Wavelength Repeatability: SD 10 meas. < 0.1 nm
Absorbance Accuracy: < ± 0.010 A at 1 A, < ± 0.010 A at 2 A
Absorbance Repeatability: ± 0.0025 A @ 1 A
Stray Light:
Photometric Accuracy 1 A: ± 0.004 A, 2 A: ± 0.004 A, 3 A: ± 0.006 A, Measured at 440 nm using neutral density filters traceable to NIST
Photometric Display ± 6 A
Photometric Range > 4 A
Photometric Repeatability 1 A: ± 0.0025 A
Scan Ordinate Modes Absorbance, % Transmittance, % Reflectance, Kubelka-Munk, Log(1/R), Log(Abs), ABS × Factor, Intensity, first to fourth Derivative
Scan Speed Variable, up to 6000 nm/min.
Stray Light KCl, 198 nm: ≤ 0.4 % T
NaI, 220 nm: ≤ 0.032 % T
NaNO2, 340 nm: < 0.01 % T
Warranty Three-year Source Replacement Warranty
Wavelength Accuracy ± 0.20 nm (546.07 nm Hg emission line), ± 0.30 nm, 190-900 nm
Wavelength Data Interval 10, 5, 2, 1, 0.5, 0.2, 0.1, 0.05 nm
Wavelength Repeatability Standard deviation of 10 measurements < 0.05 nm
Detector Type Dual Matched Silicon
Photodiodes
Light Source Xenon Flash
Product Line Evolution Pro
Spectral Bandwidth 0.5, 1, 1.5, 2, 4 nm Selectable
Type UV-Vis Spectrophotometer
Wavelength Range 190-1100 nm
Unit Size Each

Other Equipment by this Supplier

Other Equipment

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.

Your questions, but not your email details will be shared with OpenAI and retained for 30 days in accordance with their privacy principles.

Please do not ask questions that use sensitive or confidential information.

Read the full Terms & Conditions.