The Thermo Scientific™ Evolution™ Pro UV-Vis Spectrophotometer line integrates durable, compatible hardware with a broad selection of accessories, and practical, easy-to-use Thermo Scientific™ Insight™ Pro Software. The result is routine, complex research and advanced applications, including current compliance.
Product Overview
- Workflow-focused application modules make it easier to customize even the most intricate processes.
- Users may swiftly move from sample to final report, thanks to the user-friendly INSIGHT Pro Software, which supports every step of the workflow.
- Users can achieve the findings they need for quantitative analysis, scanning, and kinetics applications, with extensive software capabilities for data collection, processing, and reporting.
- By guaranteeing data security, optional INSIGHT Pro Security Software capabilities help pharmaceutical laboratories comply with current 21 CFR Part 11 regulations.
Enhance Productivity with Sampling Flexibility
- Thermo Scientific™ Smart Accessories feature a snap-in design that ensures ease of use and consistency, enabling users to complete analysis more quickly.
- Make use of the broad beam separation and big sample compartment of the Evolution Pro UV-Vis Spectrophotometer, which enables unique accessories and creative applications.
- Manual setup is eliminated by software integration and automatic recognition of sippers, cell changers, and autosampler accessories.
Regulatory/Quality Control Environments
To help users use the Evolution Pro UV-Vis Spectrophotometer in regulated or quality-control environments while reducing installation time and preserving system performance, a variety of compliance solutions and services are available.
Evolution Pro Pharma Value Bundles Include
- Evolution Pro Spectrometer
- INSIGHT Pro with Security Software
- Qualification Filter Kit for UV-Vis
- Validator for Insight Pro US Ltr or INTL A4 Paper Version
Evolution Pro USP/EP Standards Pharma Bundles Include
- USP and EP Visible Standards Set
- Validator for Insight Pro US Ltr or INTL A4 Paper Version
- Evolution Pro Spectrometer
- INSIGHT Pro with Security Software
- USP and EP UV Standards Set
Accessory Ordering Information
Each accessory can be ordered individually. Cell changers, cuvettes and test holders, calibration carousels, sippers, fiber optic probes, small-volume sample instruments, and accessories for quick mixing and reflectance are further choices.
Download the Brochure for More Information
Specifications
Source: Thermo Fisher Scientific - UV-Vis Spectroscopy
| Specifications |
| Baseline Flatness |
± 0.001 A (200 to 800 nm), 2 nm SBW, smoothed |
| Drift |
< 0.0005 A&fraslhr., 500 nm, 2 nm SBW, 2-hour warmup |
| Electrical Requirements |
100/240 V, 50/60 Hz |
| For Use With (Application) |
Academic, Chemicals, Food and Beverage, Chemicals Environmental, Industrial QA/QC, Material Science, Pharmaceuticals |
| Green Features |
ACT Ecolabel (Gold) |
| Includes |
Evolution Pro UV-Vis Spectrophotometer |
| Lamp Life |
Five years or longer if not using a live signal |
| Noise |
0 A: < 0.00018 A 1 A: < 0.00022 A 2 A: < 0.00050 A 500 nm, 2 nm SBW, RMS |
| Optical Design |
Modified Ebert Double beam with sample and reference cuvette/accessory positions |
Pharmacopoeia Compliance Testing |
Wavelength Accuracy: ± 0.3 nm 190–900 nm Wavelength Repeatability: SD 10 meas. < 0.1 nm Absorbance Accuracy: < ± 0.010 A at 1 A, < ± 0.010 A at 2 A Absorbance Repeatability: ± 0.0025 A @ 1 A Stray Light: |
| Photometric Accuracy |
1 A: ± 0.004 A, 2 A: ± 0.004 A, 3 A: ± 0.006 A, Measured at 440 nm using neutral density filters traceable to NIST |
| Photometric Display |
± 6 A |
| Photometric Range |
> 4 A |
| Photometric Repeatability |
1 A: ± 0.0025 A |
| Scan Ordinate Modes |
Absorbance, % Transmittance, % Reflectance, Kubelka-Munk, Log(1/R), Log(Abs), ABS × Factor, Intensity, first to fourth Derivative |
| Scan Speed |
Variable, up to 6000 nm/min. |
| Stray Light |
KCl, 198 nm: ≤ 0.4 % T NaI, 220 nm: ≤ 0.032 % T NaNO2, 340 nm: < 0.01 % T |
| Warranty |
Three-year Source Replacement Warranty |
| Wavelength Accuracy |
± 0.20 nm (546.07 nm Hg emission line), ± 0.30 nm, 190-900 nm |
| Wavelength Data Interval |
10, 5, 2, 1, 0.5, 0.2, 0.1, 0.05 nm |
| Wavelength Repeatability |
Standard deviation of 10 measurements < 0.05 nm |
| Detector Type |
Dual Matched Silicon Photodiodes |
| Light Source |
Xenon Flash |
| Product Line |
Evolution Pro |
| Spectral Bandwidth |
0.5, 1, 1.5, 2, 4 nm Selectable |
| Type |
UV-Vis Spectrophotometer |
| Wavelength Range |
190-1100 nm |
| Unit Size |
Each |