The durable Thermo Scientific™ GENESYS™ 50 UV-Visible and GENESYS 40 Vis Spectrophotometers conduct routine measurements for quality assurance or educational purposes in the chemistry classroom.
These instruments, available in a single-cell configuration, feature a sloped surface to facilitate spill management and a washable sampling compartment for easy cleaning.
The GENESYS™ 50 UV-Vis or GENESYS 40 Vis Spectrophotometer is the optimal selection for daily sampling tasks, featuring a 7-inch touchscreen interface, local control software (no PC required), and optional Wi-Fi connectivity.
The GENESYS™ 50 is a dual-beam UV-Vis spectrophotometer configured in a single-cell format that fulfills all requirements for critical quality control tasks and accurate chemical measurements.
The GENESYS 40 Vis Spectrophotometer is the visible-spectral-region variant in this category of instruments. Both models offer data resolution ranging from 0.2 nm to 5 nm.
Features
- Spacious sample compartment with front access
- 7-inch high-definition touchscreen user interface
- Networking enabled (Wi-Fi adapter available)
- Local control ensures speed and reliability
Designed for
- Moderate throughput requirements
- Routine measurements
- Chemistry labs
Download the Brochure for More Information
Specifications
Source: Thermo Fisher Scientific - UV-Vis Spectroscopy
| . |
. |
| Baseline Flatness |
±0.005 A |
| Compartments |
Accessible from top, front, or side Able to accommodate cells up to 100mm cells Available accessory for test tubes up to 25 mm in diameter and 150 mm in height Removable, washable sample compartment liner with magnetic placement and hold-down |
| Connections |
Single USB-A Duplex USB-A Export data to network or PC via USB, Ethernet, or Wi-Fi USB adaptor, available Print via USB, Ethernet, or Wi-Fi USB adaptor available |
| Dimensions (L x W x H) |
35.5 x 38.5 x 19.5 cm |
| Drift |
(At 500 nm after 1 hour warmup) <0.0010 A/hr. |
| Electrical Requirements |
External AC to DC converter. Voltage and Frequency (Hz) selected automatically, 100/240 V, 50/60 Hz |
| Includes |
GENESYS 40 Spectrophotometer |
| Keypad |
Touchscreen |
| Lamp Life |
>1000 hrs. |
| Noise |
(RMS at 500 nm, 60 consecutive measurements) ≤0.00020A at 0A at 260 and 500 nm ≤0.00030A at 1A at 260 and 500 nm ≤0.00040A at 2A at 260 and 500 nm |
| Optical Design |
Dual Beam |
| Photometric Accuracy |
±0.002 A at 0.5 A, ±0.004 A at 1.0 A, ±0.008 A at 2.0 A |
| Photometric Display |
-3 A to +5 A |
| Photometric Range |
–3 A to +3.5 A |
| Photometric Repeatability |
±0.001 A at 1 A measured at 1.0A at 546 nm |
| Printer Included |
Snap-on printer available |
| Scan Speed |
Automatic – up to 1800 nm/min. |
| Stray Light |
<0.05 %T at 340 nm and 400 nm |
| Warranty |
2 yrs std + 1 yr ext upon reg |
| Wavelength Accuracy |
±0.5 nm |
| Wavelength Data Interval |
0.2 nm, 0.5 nm, 1 nm, 2 nm, 5 nm |
| Wavelength Repeatability |
< ±0.2 nm |
| Weight (Metric) |
7.5 kg |
| Detector Type |
Dual Silicon Photodiodes |
| Display |
7 in. High Definition Color Touchscreen, Fixed, 800 x 1280 pixels |
| Light Source |
Tungsten Halogen |
| Spectral Bandwidth |
5 nm |
| Wavelength Range |
325 to 1100 nm |
| Unit Size |
Each |
Videos
Scan Method Overview - GENESYS UV-Vis Spectrophotometer. Video Credit: Thermo Fisher Scientific - UV-Vis Spectroscopy
Quant method Overview - GENESYS UV-Vis Spectrophotometer. Video Credit: Thermo Fisher Scientific - UV-Vis Spectroscopy
Fixed Method Overview - GENESYS UV-Vis Spectrophotometer. Video Credit: Thermo Fisher Scientific - UV-Vis Spectroscopy
Easy clean up with GENESYS Spectrophotometer washable sample tray. Video Credit: Thermo Fisher Scientific - UV-Vis Spectroscopy
Get to work quickly with GENESYS 50 Spectrophotometers. Video Credit: Thermo Fisher Scientific - UV-Vis Spectroscopy
Improving the quality of bacterial culture and oligonucleotides with UV-Vis spectroscopy to accelerate mRNA vaccine production. Video Credit: Thermo Fisher Scientific - UV-Vis Spectroscopy