The automated and network-capable Thermo Scientific™ GENESYS™ 150 UV-Visible Spectrophotometer can be used for quantitative measurements in standard industrial QA/QC or university chemistry labs.
The spectrophotometer's performance and usability have been improved with a high-resolution color touchscreen, a robust construction suited for intensive, repetitive use, and optional WiFi networking.
The GENESYS spectrophotometers, renowned worldwide for their precision, dependability, and repeatability, meet modern demands for cutting-edge technology in a small, robust package.
The GENESYS 150 UV/Vis Spectrophotometer offers lid-open operation for maximum speed and convenience, and provides automation for high-throughput applications and room-light resistance.
To simplify sampling and enable high-throughput measurements and temperature control, a range of attachments has been developed, including automated cell changers and cell holders, fiber-optic probes, and sippers. Data resolution ranges from 0.2 to 5 nm.
Among the instrumental options, the GENESYS 140 Vis Spectrophotometer is a visible choice. The data resolution of both devices ranges from 0.2 to 5 nm.
Key Features
- It is simple to remove and wash the sample container liner, thanks to its magnetic positioning and hold-down.
- Peltier-thermostatted single cell holder (20 - 60 °C) (optional)
- 8-position cell changer with switchable turret (optional)
- 4-position cell changer (long path cells) (optional)
- Contains cells with a path length of up to 100 mm
- Networking capable (with optional WiFi adaptor)
- Onboard techniques, reporting, and data export
- Room light resistance for “open-lid” operation
- 7″ high-resolution touchscreen user interface
- Sipper accessory (optional)
Designed for
- Applications requiring high throughput or temperature control
- Research and development departments
- Sophisticated teaching labs
Download the Brochure for More Information
Specifications
Source: Thermo Fisher Scientific - UV-Vis Spectroscopy
| . |
. |
| Baseline Flatness |
±0.005 A |
| Compartments |
Accommodates (optional): 8-position cell changer 4-position cell changer (long path cells) Peltier thermostatted cell holder (20-60 °C) Sipper accessory Fiber optic probe coupler |
| Connections |
Single USB-A Duplex USB-A Export data to network or PC via USB, Ethernet, or Wi-Fi USB adaptor, available Print via USB, Ethernet, or WiFi. USB adaptor available |
| Dimensions (L x W x H) |
35.5 x 38.5 x 19.5 cm |
| Drift |
(At 500 nm after 1 hour warmup) <0.0010 A/hr. |
| Electrical Requirements |
External AC to DC converter. Voltage and Frequency (Hz) selected automatically, 100/240 V, 50/60 Hz |
| Includes |
Genesys 140 Spectrophotometer |
| Keypad |
Touchscreen |
| Lamp Life |
>1000 hrs. |
| Noise |
(RMS at 500 nm, 60 consecutive measurements) ≤0.00020 A at 0 A at 260 and 500 nm ≤0.00030 A at 1 A at 260 and 500 nm ≤0.00040 A at 2 A at 260 and 500 nm |
| Optical Design |
Dual Beam |
| Photometric Accuracy |
±0.002 A at 0.5 A, ±0.004 A at 1.0 A ±0.008 A at 2.0 A |
| Photometric Display |
-3 A to +5 A |
| Photometric Range |
-3 A to +3.5 A |
| Photometric Repeatability |
±0.001 A at 1 A measured at 1.0 A at 546 nm |
| Printer Included |
Snap-on printer available |
| Scan Speed |
Automatic – up to 1800 nm/min. |
| Stray Light |
<0.05 %T at 340 nm and 400 nm |
| Warranty |
2 yrs standard + 1 year extension upon regulation |
| Wavelength Accuracy |
±0.5 nm |
| Wavelength Data Interval |
0.2 nm, 0.5 nm, 1 nm, 2 nm, 5 nm |
| Wavelength Repeatability |
< ±0.2 nm |
| Weight (Metric) |
7.5 kg |
| Detector Type |
Dual Silicon Photodiodes |
| Display |
7 in. High Definition Color Touchscreen, Tiltable, 800 x 1280 pixels |
| Light Source |
Tungsten Halogen |
| Spectral Bandwidth |
5 nm |
| Wavelength Range |
325 to 1100 nm |
| Unit Size |
Each |