GENESYS™ 180 Double Beam UV-Vis Spectrophotometer

The Thermo Scientific GENESYS 180 UV-Visible Spectrophotometer enables high-throughput quantitative UV-Vis readings using a double-beam reference cell. The GENESYS 180 UV-Visible Spectrophotometer's high-resolution color touchscreen, optional WiFi networking, and sturdy build for use in demanding, repetitive settings have been included to maximize usability and performance.

The precision, reliability, and reproducibility of GENESYS spectrophotometers are recognized worldwide. As a result, these spectrophotometers meet modern demands for advanced technology in a compact, reliable form.

For labs that require a reference cell location for applications such as kinetics, the GENESYS 180 UV-Vis Spectrophotometer provides double-beam operation. For high-throughput applications, an 8-cell carousel is the standard. The spectrophotometer's resistance to ambient light enables lid-open operation for greater convenience.

Fiber optic probes, automatic cell changers and holders, and a sipper are just a few of the accessories that come with the GENESYS 180 UV-Vis Spectrophotometer. Measurements outside the sample chamber are made easier with these devices. They are designed to facilitate high-throughput measurements, ensure temperature control, and simplify sampling.

Features

  • Peltier-thermostatted single cell holder (20 - 60 °C) (optional)
  • Standard 8-position cell changer including switchable turret
  • Long-path cells with a four-position cell changer (optional) 
  • Cells with a path length of up to 100 mm are supported
  • Networking capability (with optional Wi-Fi adaptor)
  • 7-inch high-resolution touchscreen user interface
  • Onboard methods, reporting, and data export 
  • Room light resistance for “open-lid” operation
  • Double-beam/reference cell
  • Sipper accessory (optional)

Download the Brochure for More Information

Specifications

Source: Thermo Fisher Scientific - UV-Vis Spectroscopy 

. .
Baseline Flatness ±0.002 A
Compartments 8-position cell changer (standard)
Also accommodates (optional):
4-position cell changer (long path cells)
Peltier thermostatted cell holder (20-60 °C)
Sipper accessory
Fiber optic probe coupler
Connections Single USB-A
Duplex USB-A
Export data to network or PC via USB, Ethernet, or Wi-Fi USB adaptor, available
Print via USB, Ethernet, or Wi-Fi
USB adaptor available
Dimensions (L x W x H) 35.5 x 38.5 x 19.5 cm
Drift (At 500 nm after 1 hour warmup) <0.0005 A/hr.
Electrical Requirements External AC to DC converter.
Voltage and Frequency (Hz) selected automatically, 100/240 V, 50/60 Hz
Keypad Touchscreen
Lamp Life > 5 years typical, 3 years guaranteed
Noise (RMS at 500 nm, 60 consecutive measurements)
≤0.00020 A at 0 A at 260 and 500 nm
≤0.00030 A at 1 A at 260 and 500 nm
≤0.00040 A at 2 A at 260 and 500 nm
Optical Design Double Beam
Photometric Accuracy ±0.002 A at 0.5 A, ±0.004 A at 1.0 A, ±0.008 A at 2.0 A
Photometric Display -3 A to +5 A
Photometric Range –2 A to +3.5 A
Photometric Repeatability ±0.001 A at 1 A measured at 1.0 A at 546 nm
Printer Included Snap-on printer available
Scan Speed Slow, medium, and fast - up to 1600 nm/min.
Stray Light < 1.0 %T 198 nm (KCl), <0.05 %T at 220 nm (NaI), <0.03 %T at 340 nm (NaNO2)
Warranty 2 yrs std + 1 yr ext upon reg
Wavelength Accuracy ±0.5 nm
Wavelength Data Interval 0.1 nm, 0.2 nm, 0.5 nm, 1 nm, 2 nm, 5 nm
Wavelength Repeatability < ±0.2 nm
Weight (Metric) 7.5 kg
Detector Type Dual Silicon Photodiodes
Display 7 in. High Definition Color Touchscreen, Tiltable, 800 x 1280 pixels
Spectral Bandwidth 2 nm
Type UV-Vis Spectrophotometer
Wavelength Range 190 to 1100 nm
Unit Size Each

Other Equipment by this Supplier

Other Equipment

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.

Your questions, but not your email details will be shared with OpenAI and retained for 30 days in accordance with their privacy principles.

Please do not ask questions that use sensitive or confidential information.

Read the full Terms & Conditions.