Posted in | Gas Analyzers

Multi-Gas Calibrator for EPA-Compliant Ambient Air Monitoring

The Thermo Scientific iQ Series Gas Analyzer provides 24/7 access to analyzer information. The iQ Series Gas Analyzers are designed to get smarter over time, reducing troubleshooting time with a highly precise but robust solution.

When calibrating ambient air and other gas monitoring equipment, the Thermo Scientific 146iQ Multi-Gas Calibrator offers the same level of performance. In compliance with US EPA regulations, devices performing zero, precision, Level 1 span checks, audits, and multi-point procedures are calibrated using the output of the 146iQ Multi-Gas Calibrator.

Discover the nonstop intelligence of the iQ Series Gas Analyzer, now with a two-year guarantee. This next generation platform for gas analyzers includes:

Predictive Diagnostics

With better serviceability and plug-and-play components, issues may be resolved before they arise. With creative, replaceable DMC modules, troubleshooting time is reduced with round-the-clock access to the analyzer's data in a highly accurate but durable solution.

Proactive Communications

Maintain peak performance using iQ360 and email notifications. When users subscribe to the iQ360 service, the instrument sends an alert straight to the technical support staf, who then create a remedial action plan and send it to the staff via email. The iQ Series Gas Analyzers eliminate surprises and give users complete control.

Personal Device Connectivity

The iQ companion mobile device app now enables continuous, real-time remote monitoring, which makes it easier to contact Thermo Scientific, enables remote monitoring of iQ gas analyzers, and provides immediate access to product documentation and maintenance notifications.

Specifications

Source: Thermo Fisher Scientific – Environmental and Process Monitoring Instruments

Specifications
Accuracy ± 0.8 % of reading + 0.2 % of full-scale
Certifications/Compliance CE, TUV-SUD Safety, UKCA
Communication Ports Serial ports:
One RS-232/485 port
One RS-485 external accessory port
Other ports:
Three Full Speed USB ports (one in front, two in rear)
One Gigabit ethernet port

Communication: MODBUS, Streaming, Gesytec (Bayern-Hessen)
Depth (Imperial) 24 in.
Depth (Metric) 609 mm
Description 146iQ Multi-Gas Calibrator
Electrical Requirements 100-240 VAC 50/60 Hz, 275 watts
Height (Imperial) 8.72 in.
Height (Metric) 221.48 mm
Outputs Six isolated voltages outputs
Six isolated analog current outputs
Precision 1 ppb
Repeatability of Flow Control ± 0.2 % of full scale
Span Drift < 1 % full scale (one month)
Temperature (Imperial)
Operating
32 °F to 113 °F
Voltage 100/240 V
Weight (Imperial) 43 lb. (with three MFCs, Ozonator, and Photometer)
Weight (Metric) 19.5 kg (with three MFCs, Ozonator, and Photometer)
Width (English) 16.75 in.
Width (Metric) 425.45 mm
Zero Drift < 1.0 ppb (24 hours)
Detection Limits 1.5 ppb (60 second averaging time)
Flow Rate 0.8 L/min.
Inputs 16 Inputs (TLL), four Isolated
Voltage Inputs 0–10 V
Linearity ± 0.25 % full scale
Response Time 20 seconds (10 second averaging) (Photometer),
< 180 sec. (Ozonator Response Time)
Temperature (Metric)
Operating
0–45 °C
Zero Noise 0.75 ppb RMS (60 second averaging time)
Unit Size Each

Other Equipment by this Supplier

Other Equipment

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.

Your questions, but not your email details will be shared with OpenAI and retained for 30 days in accordance with their privacy principles.

Please do not ask questions that use sensitive or confidential information.

Read the full Terms & Conditions.