The Thermo Scientific™ Personal DataRAM pDR-1500 Portable Monitor is a fully integrated, high-precision, real-time piece of sampling equipment that offers optimal ease-of-use and enhanced operating time.
The pDR-1500 has the capability and versatility to offer real-time findings and gravimetric validation for a variety of applications, including site remediation monitoring, health impacts studies, and construction/demolition projects. At 41 ounces, it is just as convenient to wear on the belt as it is to run in a fixed position.
The pDR-1500 is equipped to compensate for all usual environmental factors, including relative humidity, real volumetric flow, pressure compensation, and historical pDR nephelometry technology. The pDR-1500 aerosol monitor additionally has replaceable cyclones for improved particle-cut points (PM10 & PM4 or PM2.5 & PM1), as well as an integrated sample filter that allows for post-gravimetric validation of all data.
Key Features
- Accurate volumetric flow control
- Complete compensation for environmental variables
- Flexible data logging routine
Additional Information
- Number of data tags: 99 (maximum)
- Serial interface: USB/RS-232, 19,200 baud
- Computer requirements: PC compatible, 486 processor or higher; Windows™ 95 or higher
- Aerodynamic particle cut-point ange 1–10 μm
- Concentration display updating interval : one second
- Data logging averaging periods (user selectable): one second to one hour
- Total number of data points in memory: greater than 500,000
Recommended for
- Exposure modeling
- Protection of asthma patients
- Site remediation
- Size discrimination
- Mass validation
Specifications
Source: Thermo Fisher Scientific – Environmental and Process Monitoring Instruments
| Specifications |
| Accuracy |
± 5 % of reading ± precision (traceable to SAE Fine Test Dust) |
Aerodynamic Particle Cut-Point Range |
1–10 μm |
| Certifications/ Compliance |
Suitable for NIOSH Methods 0500 and 0600. CE marked. UKCA |
Concentration Display Averaging Time |
One to 60 seconds (user selectable) |
Concentration Display Updating Interval |
One second |
| Concentration Measurement Range |
0.001–400 mg/m3 range (auto ranging) |
| Data Points |
> 500,000 |
| Depth (Imperial) |
3.3 in. |
| Depth (Metric) |
84 mm |
| Description |
personalDataRAM pDR-1500 Aerosol Monitor |
| Height (Imperial) |
7.1 in. |
| Height (Metric) |
181 mm |
| Particle Size Range |
0.1–10 µm |
| Precision |
± 2 % of reading or ± 0.005 mg/m3, whichever is larger, for one-second (2-sigma)2 averaging time, ± 0.5 % of reading or ± 0.0015 mg/m3, whichever is larger, for 10-second averaging time, ± 0.2 % of reading or ± 0.0005 mg/m3, whichever is larger, for 60-second averaging time |
| Resolution |
0.1 % of reading or 0.001 mg/m3, whichever is larger |
| Scattering Coefficient Range |
1.5 x 10-6 to 0.6 m-1 (approx.) @ lambda= 880 nm (not displayed) |
| Weight (Imperial) |
41 oz. |
| Weight (Metric) |
1.2 kg |
| Width (Imperial) |
5.6 in. |
| Width (Metric) |
143 mm |
| Flow Rate |
1–3.5 L/min. |
| Unit Size |
Each |